User Contributed MET/CAL PROCEDURE ============================================================================= INSTRUMENT: Elgar 403-VP:Oscillator DATE: 06-Aug-97 AUTHOR: User Contributed REVISION: 0 ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 20 NUMBER OF LINES: 177 CONFIGURATION: HP 34401A CONFIGURATION: HP 8903A ============================================================================= STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 ASK- R N P F W 1.002 ASK+ X 1.003 HEAD INITIAL CONDITIONS 1.004 DISP Insert the UUT into a single phase Elgar power supply. 1.005 DISP Plug the Elgar power supply into 208 volt 3 phase power 1.005 DISP receptacle. 1.006 DISP Connect the line to neutral output from the Elgar 1.006 DISP to the HI/LO INPUT on the 34401A. 1.007 DISP Turn on the Elgar power supply. 1.008 DISP Set the UUT contols as follows: 1.008 DISP FREQ. MULTIPLIER ............. X1 1.008 DISP VERNIER ...................... CAL 1.008 DISP frequency dials .............. (10)00 1.009 IEEE [@34401]FUNC "VOLT:AC"[GTL] 1.010 DISP Adjust the Elgar output for a reading of 20 volts 1.010 DISP on the 34401A. 1.011 HEAD {} 1.012 HEAD {* Multiplier Accuracy Test *} 1.013 JMP 2.001 1.014 EVAL 2.001 DISP Set the UUT contols as follows: 2.001 DISP FREQ. MULTIPLIER ............. X1/10 2.001 DISP frequency dials .............. 100 2.002 IEEE [@34401]FUNC "FREQ" 2.003 IEEE [@34401]FREQ:APER MAX 2.004 ACC 10.000H 0.1% 2.005 IEEE [@34401][D2000]READ?[T0][I] 2.006 MEME 2.007 MEMC H 10% X1/10 3.001 DISP Set the UUT contols as follows: 3.001 DISP FREQ. MULTIPLIER ............. X1 3.002 IEEE [@34401]FUNC "FREQ" 3.003 IEEE [@34401]FREQ:APER MAX 3.004 ACC 100.00H 0.1% 3.005 IEEE [@34401][D2000]READ?[T0][I] 3.006 MEME 3.007 MEMC H 10% X1 4.001 DISP Set the UUT contols as follows: 4.001 DISP FREQ. MULTIPLIER ............. X10 4.002 IEEE [@34401]FUNC "FREQ" 4.003 IEEE [@34401]FREQ:APER MAX 4.004 ACC 1000.0H 0.1% 4.005 IEEE [@34401][D2000]READ?[T0][I] 4.006 MEME 4.007 MEMC H 10% X10 5.001 HEAD {} 5.002 HEAD {* Dial Linearity Test *} 5.003 JMP 6.001 5.004 EVAL 6.001 DISP Set the UUT contols as follows: 6.001 DISP FREQ. MULTIPLIER ............. X1 6.001 DISP frequency dials .............. 111 6.002 IEEE [@34401]FUNC "FREQ" 6.003 IEEE [@34401]FREQ:APER MAX 6.004 ACC 111.00H 0.1% 6.005 IEEE [@34401][D2000]READ?[T0][I] 6.006 MEME 6.007 MEMC H 10% 7.001 DISP Set the UUT contols as follows: 7.001 DISP frequency dials .............. 222 7.002 IEEE [@34401]FUNC "FREQ" 7.003 IEEE [@34401]FREQ:APER MAX 7.004 ACC 222.00H 0.1% 7.005 IEEE [@34401][D2000]READ?[T0][I] 7.006 MEME 7.007 MEMC H 10% 8.001 DISP Set the UUT contols as follows: 8.001 DISP frequency dials .............. 333 8.002 IEEE [@34401]FUNC "FREQ" 8.003 IEEE [@34401]FREQ:APER MAX 8.004 ACC 333.00H 0.1% 8.005 IEEE [@34401][D2000]READ?[T0][I] 8.006 MEME 8.007 MEMC H 10% 9.001 DISP Set the UUT contols as follows: 9.001 DISP frequency dials .............. 444 9.002 IEEE [@34401]FUNC "FREQ" 9.003 IEEE [@34401]FREQ:APER MAX 9.004 ACC 444.00H 0.1% 9.005 IEEE [@34401][D2000]READ?[T0][I] 9.006 MEME 9.007 MEMC H 10% 10.001 DISP Set the UUT contols as follows: 10.001 DISP frequency dials .............. 555 10.002 IEEE [@34401]FUNC "FREQ" 10.003 IEEE [@34401]FREQ:APER MAX 10.004 ACC 555.00H 0.1% 10.005 IEEE [@34401][D2000]READ?[T0][I] 10.006 MEME 10.007 MEMC H 10% 11.001 DISP Set the UUT contols as follows: 11.001 DISP frequency dials .............. 666 11.002 IEEE [@34401]FUNC "FREQ" 11.003 IEEE [@34401]FREQ:APER MAX 11.004 ACC 666.00H 0.1% 11.005 IEEE [@34401][D2000]READ?[T0][I] 11.006 MEME 11.007 MEMC H 10% 12.001 DISP Set the UUT contols as follows: 12.001 DISP frequency dials .............. 777 12.002 IEEE [@34401]FUNC "FREQ" 12.003 IEEE [@34401]FREQ:APER MAX 12.004 ACC 777.00H 0.1% 12.005 IEEE [@34401][D2000]READ?[T0][I] 12.006 MEME 12.007 MEMC H 10% 13.001 DISP Set the UUT contols as follows: 13.001 DISP frequency dials .............. 888 13.002 IEEE [@34401]FUNC "FREQ" 13.003 IEEE [@34401]FREQ:APER MAX 13.004 ACC 888.00H 0.1% 13.005 IEEE [@34401][D2000]READ?[T0][I] 13.006 MEME 13.007 MEMC H 10% 14.001 DISP Set the UUT contols as follows: 14.001 DISP frequency dials .............. 999 14.002 IEEE [@34401]FUNC "FREQ" 14.003 IEEE [@34401]FREQ:APER MAX 14.004 ACC 999.00H 0.1% 14.005 IEEE [@34401][D2000]READ?[T0][I] 14.006 MEME 14.007 MEMC H 10% 15.001 DISP Set the UUT contols as follows: 15.001 DISP frequency dials .............. (10)99 15.002 IEEE [@34401]FUNC "FREQ" 15.003 IEEE [@34401]FREQ:APER MAX 15.004 ACC 1100.00H 0.1% 15.005 IEEE [@34401][D2000]READ?[T0][I] 15.006 MEME 15.007 MEMC H 10% 16.001 HEAD {} 16.002 HEAD {* Vernier Accuracy Test *} 16.003 JMP 17.001 16.004 EVAL 17.001 DISP Set the UUT contols as follows: 17.001 DISP frequency dials .............. (10)00 17.001 DISP VERNIER ...................... -2% 17.002 IEEE [@34401]FUNC "FREQ" 17.003 IEEE [@34401]FREQ:APER MAX 17.004 ACC 980.00H 0.1% 17.005 IEEE [@34401][D2000]READ?[T0][I] 17.006 MEME 17.007 MEMC H 10% 18.001 HEAD {} 18.002 HEAD {* Distortion Test *} 18.003 JMP 19.001 18.004 EVAL 19.001 DISP Set the Elgar output to zero. 19.002 DISP Disconnect the Elgar from the 34401A and connect it 19.002 DISP to the HP 8903A Distortion Analyzer. 19.003 IEEE [@8903][TERM LF] 19.004 IEEE [@8903]M1H0 19.005 DISP Adjust the Elgar output for a reading of 20 volts 19.005 DISP on the 8903A. 19.006 ACC 1pct 0.015U 19.007 IEEE [@8903]M3LN 19.008 IEEE [@8903][I] 19.009 MATH mem1=0 19.010 MEME 19.011 MEMC pct +1.00U 1000H 20.001 HEAD {} 20.002 END